商品規格 |
商品名稱:
|
總硫分析儀 TCL-100型
Trace Chlorine/Sulfur Analyzer model TCL-100
《垂直爐燃燒方式》
|
分析方法:
|
高溫氧化分解庫倫滴定法 |
燃燒:
|
矽管燃燒
|
高溫爐:
|
1100oC |
樣品種類:
|
液體及氣體樣品 |
檢測方法:
|
氧化-還原電位(以電極檢測) |
滴定控制:
|
自動電解電流控制 |
測定範圍:
|
微量~1000ppm |
測定時間:
|
約10分 |
操作控制:
|
PC配合軟體程式Data Base |
所需氣體:
|
Ar 99.8%以上, O2 99.7%以上 |
符合標準:
|
氯:ASTM D5808, ASTM D4929…等 |
|
|
|
|
TCL-100 |
The field-proven coulometry system allows quick measurement of all chlorine contained in a liquid or gaseous sample in different concentrations ranging form ppm to percent levels. Our analyzer allows fully automatic measurement of all chlorine in waste oil or reclaimed oil to reduce petroleum products, petrochemical products, and dioxin.
|
Features
Windows® XP is employed for easier operation and advanced functions.
You can easily operate the system with mouse functions. The heater and the coulometer can be switched on or off and gas flow rate monitor can be controlled with a PC. In addition, when the system is in operation, the monitor provides a view of the reaction condition and the progress of measurement.
Easy automatic measurement
Connecting the liquid auto sampler (ASC-150L) allows automatic measurement of samples. This makes your research, routine analysis, or quality control analysis more efficient.
Allows measurement of a trace quantity of sulfur/chlorine
By just replacing the chlorine/sulfur titration cell with the optional sulfur/Chlorine titration cell, you can measure all sulfur/Chlorine contained in different concentrations from ppm to percent levels. Optional products for common use with chlorine/sulfur are available.
Configurable into a desired system
There are different feeders available for different samples. You can select the appropriate ones for the samples to analyze and configure a desired system.
Employed using different standard test methods
- [Chlorine analyzer]
- ASTM D5808:
- Standard Test Method for Determining Organic Chloride in Aromatic Hydrocarbons and Related Chemicals by Microcoulometry
- ASTM D4929:
- Standard Test Methods for Determination of Organic chloride content in Crude Oil.
- [Sulfur analyzer]
- JIS K2541:
- Crude oil and petroleum products- Determination of sulfur content
- JIS K2421:
- Methods for testing benzenes
- JIS K2240:
- Liquefied petroleum gases
- JPI-55-20-82:
- Light oil and petroleum products- Determination of sulfur content
- ASTM D3120:
- Standard Test Method for Trace Quantities of Sulfur in Light Liquid Petroleum Hydrocarbons by Oxidative Microcoulometry
- ASTM D3140:
- Standard Test Method for Sulfur in Petroleum Gas by Oxidative Microcoulometry
|
TCL-100 Specifications |
|
Analysis method: |
Oxidative decomposition, coulometric titration |
|
Combustion: |
Silica tube combustion |
|
Combustion furnace
temperature: |
1,100oC at he maximum |
|
Samples to analyze: |
Liquid and gaseous samples |
|
Detection method: |
Oxidation-reduction potential
(electric potential detection by electrodes) |
|
Detecting electrode: |
[Chlorine analyzer]: Silver electrode
[Sulfur analyzer]: Alloy electrode |
|
Titration control: |
Automatic electrolytic current control |
|
Gas to use: |
Argon: Purity greater than 99.98%, 300±100kPa
Oxygen: Purity greater than 99.7%, 300±100kPa |
|
Measurement range:
[Chlorine analyzer] |
0.1 to 50μg (CI), 0.5 to 1,000ppm
(Samples burnable below 1,000oC with no difficulty) |
|
Measurement accuracy:
[Chlorine analyzer] |
Standard samples: CV of 5% or below for 1μg of chlorine
When 2,4-dinitrochlorobenzene is measured |
|
Measurement range:
[Sulfur analyzer] |
0.02 to 20μg (S), 0.1 to 1,.000ppm
(Samples burnable below 1,000oC with no difficulty) |
|
Measurement accuracy:
[Sulfur analyzer] |
Standard samples: CV of 5% or below for 1μg of sulfur
When dibutyl disulfide is measured |
|
Measurement time: |
approx. 10min./sample (content of 2μg at measurement) |
|
Amount of samples
to measure: |
Liquid: 5 to 200μl*1, Gas: 5to 10ml*2 |
|
Dimensions, weight: |
Main unit: approx. 500(W)×400(D)×500(H)mm, 35kg
Detector: approx.220(W)×370(D)×415(H)mm, 8kg |
|
*1 : The constant rate injector can be used with a liquid sample.
In this case, up to 200μl of a sample can be injected.
*2 : The gas injector is required for a gaseous sample.
|
|
Options
Depending on the samples to analyzed, optional products can be selected to configure a desired system. |
|
Liquid Auto Sampler
model ASC-150L
ACS-150L allows fully automatic measurement of up to 50 specimens. It supports many different control parameters to work with a wide variety of samples.
Sample: |
Liquid Sample |
Syringe: |
Microsyringe (Standard: 50μl)
Optional gas-tight syringes with a capacity of 25 and 100μl are also available. |
Accuracy: |
RSD of 2% or below (for injection of 40μl) |
Injection rate: |
Low: 0.4 to 2.0 μl/sec., High: 15 to 50 μl/sec. |
Suction rate: |
2, 5, 10, 20 μl/sec. |
Sample vial: |
Up to 50 glass vials with a capacity of 5ml can be installed. |
Vial for washing: |
2 glass vials with a capacity of 28ml |
Washing of syringes: |
0 to 5 times (samples and solvents can be set separately.) |
Dimensions, weight: |
approx. 460(W)×320(D)×450(H)mm, 16kg |
|
|
Constant Rate Injector
model CRI-100V
CRI-100V can inject non-aqueous samples at a constant rate to a microsyringe.
Syringe: 50μl, 100μl, 250μl (gas-tight syringes)
Injection rate: 0.2 to 1.3μl/sec. (variable)
Dimensions, weight: approx. 250(W)×150(D)×288(H)mm, 7kg
|
|
Gas Injector
model GI-100
GI-100 is used to analyze gas samples.
Samples to inject: Gaseous samples
Standard liquid sample: Max. 20μl (Gastight syringe)
Standard gaseous sample: Max. 10μl (Gastight syringe)
Gas to use: Argon (purity greater than 99.98%)
Dimensions, weight: approx. 120(W)×250(D)×235(H)mm, 4kg
|
|
|
|
|